Original Research

Repeatability and reproducibility of horizontal corneal diameter and anterior corneal power measurements using the Oculus Keratograph 4

Khathutshelo P. Mashige
African Vision and Eye Health | Vol 75, No 1 | a350 | DOI: https://doi.org/10.4102/aveh.v75i1.350 | © 2016 Khathutshelo P. Mashige | This work is licensed under CC Attribution 4.0
Submitted: 25 January 2016 | Published: 19 August 2016

About the author(s)

Khathutshelo P. Mashige, Discipline of Optometry, University of KwaZulu-Natal, South Africa

Abstract

Purpose: To evaluate the repeatability and reproducibility of horizontal corneal diameter (HCD) and anterior corneal power (ACP) measurements obtained with the Oculus Keratograph 4 (OCULUS Optikgeräte GmbH).

Methods: These parameters (HCD and ACP) were prospectively measured in quick succession three times in each of the right eyes of 40 healthy subjects, aged 18–28 years, with normal vision (6/6 or better visual acuity) in the first session by a single examiner. Measurements were then repeated in the second session scheduled 1 week later by the same examiner using the same instrument. Repeatability and reproducibility of HCD and ACP measurements was assessed based on the intra-session and intersession within-subject standard deviation (sw), repeatability (2.77sw), coefficient of variation (CoV) and intra-class correlation coefficient (ICC).

Results: Intra-session repeatability and intersession reproducibility of all measured parameters showed a repeatability (2.77sw) of 0.35 mm or less for HCD and 0.35 D or less for ACP, a CoV of 0.30% or less and an ICC of more than 0.9.

Conclusion: HCD and ACP measurements obtained using an Oculus Keratograph 4 show good repeatability and reproducibility in healthy eyes; therefore, these parameters can be used for longitudinal follow-up when measured with this device.


Keywords

Oculus Keratograph 4; repeatability; reproducibility; horizontal corneal diameter; anterior corneal power

Metrics

Total abstract views: 5455
Total article views: 9023

 

Crossref Citations

1. Normative corneal biometric parameters in African populations: A systematic review and meta-analysis
Ngozika N. Ezinne, Samuel Kyei, Chukwuemeka Junior Obinwanne, Tuwani A. Rasengane, Khathutshelo Percy Mashige, Kingsley E. Agho, Randy Asiamah, Bernadine N. Ekpenyong, Stephen Ocansey, Antor O. Ndep, Sylvester Kyeremeh, Godwin Ovenseri-Ogbomo, Edgar Ekure, Kelechi C. Ogbuehi, Kovin Shunmugan Naidoo, Nalova Westbrook, Haile Waretow, Kofi Asiedu, Uchechukwu L. Osuagwu
Contact Lens and Anterior Eye  vol: 48  issue: 4  first page: 102412  year: 2025  
doi: 10.1016/j.clae.2025.102412